Study on the Feasibility of Low-Dose 64-Slice Spiral Scanning for Head and Neck Digital Subtraction CT Angiography
cai wu,gong jianping,qian minghui,qiao fang,zhu jiangtao,zhang wei,zhang bo,dou xin,shen junkang
DOI: https://doi.org/10.13437/j.cnki.jcr.2014.07.029
2014-01-01
Abstract:Objective To investigate the clinical feasibility of low-dose 64-slice spiral CT scanning for head and neck digital subtraction CT angiography( DSCTA) by combination use of automatic tube current modulation( ATCM) technique with the different noise index( NI). Methods Ninety patients with clinically-suspected head and neck vascular diseases underwent DSCTA examination,and the patients were randomly and equally divided into 3 groups with 30 patients in each group. 64-slice spiral CT scanning was performed in all patients by combination use of ATCM technique with different NI( group A,NI =6.0 HU; group B,NI =8.0 HU and group C,NI =10.0 HU),while the other scan parameters were heldconstant. Results In group A,the mean mAs value was( 123. 03 ± 6. 17) mAs,the CTDIvolvalue was( 46. 32 ± 3. 10)mCy,the DLP value was( 1703. 40 ± 176. 70) mGy·cm and the ED value was( 3. 92 ± 0. 41) mSv,and all the values of above parameters of group A were the highest in the three groups. The results in group B were as follows: mean mAs value=( 92. 78 ± 5. 41) mAs,CTDIvolvalue =( 36. 13 ± 1. 95) mCy,DLP value =( 1340. 09 ± 122. 59) mGy·cm and the ED value =( 3. 08 ± 0. 28) mSv,the values being at the middle level among the three groups. The results in group C were as follows: mean mAs value =( 83. 08 ± 9. 25) mAs,the CTDIvolvalue =( 32. 56 ± 3. 26) mCy,the DLP value =( 1199. 86± 135. 93) mGy·cm and the ED value =( 2. 76 ± 0. 31) mSv,the above values being the lowest in the three groups( P =0. 000). For the respect of radiation dosage,group C showed a reduction of 30% and 10% when compared with group A and group B respectively,and group B showed a reduction of 21% when compared with group A. The image quality score of group A,B and C was( 4. 20 ± 0. 66),( 4. 03 ± 0. 67) and( 3. 83 ± 0. 75) respectively,the differences between each other among the three groups were not statistically significant( P > 0. 05). The mean mAs value at the head level in group A,B and C was( 104. 77 ± 12. 96) mAs,( 83. 03 ± 6. 86) mAs and( 70. 56 ± 5. 72) mAs respectively,with the value of group A being the largest and group C the smallest( P < 0. 05 in all). The image noise score at the head level in group A,B and C was( 8. 16 ± 1. 63),( 9. 63 ± 1. 03) and( 10. 54 ± 1. 72) respectively,with the image noise score of group A being the smallest and group C the largest( P < 0. 05 in all). The mean mAs value at the neck level in group A,B and C was( 99.32 ±9.80) mAs,( 62.07 ±4.46) mAs and( 58.84 ±5.47) mAs respectively,with the value of group A being the largest and group C the smallest( P < 0. 05 in all). The image noise score at the neck level in group A,B and C was( 4.93 ±0.75),( 6.16 ±0.87) and( 6.43 ±0.84) respectively,with the image noise score of group A being the smallest( P =0.000),and the difference in the image noise score between group B and C was not statistically significant( P >0. 05). The mean mAs value at the shoulder level in group A,B and C was( 147. 30 ± 1. 24) mAs,( 146. 74 ± 1. 17)mAs and( 146. 70 ± 1. 27) mAs respectively,and the image noise score at the shoulder level in group A,B and C was( 10.97 ±1.61),( 11.49 ±1.28) and( 11.61 ±1.59) respectively,and the differences between each other among the three groups were not statistically significant( P > 0. 05 in all). Conclusion It is feasible to adopt low-dose 64-slice spiral CT scanning for the performance of head and neck digital subtraction CT angiography by combination use of automatic tube current modulation technique. Scan protocol of NI = 10. 0 HU can ensure a satisfactory image quality for diagnostic purpose.