Analysis of the fast scanning method for tumor ablation with the effect of the large blood vessel by numerical simulation

Shan Qiao,Guofeng Shen,Jingfeng Bai,Yazhu Chen
DOI: https://doi.org/10.1063/1.4769961
2012-01-01
AIP Conference Proceedings
Abstract:While using HIFU for tumor ablation, the focal size of the ultrasound is relatively small compared with the tumor, therefore, numerous sonications are necessary to cover the whole treatment area. A large number of foci, on the other hand, lead to a complex problem for the optimization of the treatment parameters. Moreover, the existence of the large vessel might reduce the size of the lesion volume. A fast scanning method for volumetric ablation is investigated by numerical simulation with the effect of the large blood vessel. The proposed method is only available for phased array transducers because fast switch at the frequency of 10 Hz between several predetermined focus positions is needed. Since the duration time for each single ablation was identical, ignoring the ultrasound power, the scan path is the major parameter that should be decided. Five scan paths are simulated with and without a large vessel (diameter of 6mm) in the compute domain. The simulations solved by finite element method showed that the size of formed lesions had little difference while different scan paths applied. While the proposed scan method was used, the impact of the blood flow on the lesion volume depended on the distance between the large vessel and the focal area, as same as previous researches in the single focus case. Additionally, the orientation of the vessel played an important role in the formation of lesions.
What problem does this paper attempt to address?