Investigation of Tissue Damaging Effects of Confocal Dual-frequency High-intensity Focused Ultrasound in Transparent Phantom

KUANG Sheng-li,HU Bing,JIANG Li-xin
2008-01-01
Abstract:Objective: To evaluate the lesions size created by confocal dual-frequency high-intensity focused ultrasound (HIFU) and conventional single-frequency HIFU. Methods: Confocal dual-frequency mode and single frequency mode HIFU were used to generate ablation lesion under the same spatially averaged intensity level with different different exposure time. Optically transparent tissue mimicking phantom were used as the target of HIFU ablation. The dimensions of HIFU lesions were directly measured on optically transparent tissue mimicking phantoms. Results: The sizes of HIFU lesions enlarged gradually with the increase of exposure in both dual-frequency and single-frequency modes at the same spatially averaged intensity level,but the dimensions of lesions generated by dual-frequency mode were obviously larger than those in single-frequency mode. Conclusions: The confocal dual-frequency HIFU can enhance tissue damaging effects,representing a new technique to improve the ablation efficiency of HIFU.