Modeling and simulation of HIFU induced lesions using different treatment pattern

Shan Qiao,Guofeng Shen,Jingfeng Bai,Yazhu Chen
DOI: https://doi.org/10.1063/1.4757348
2012-01-01
AIP Conference Proceedings
Abstract:Long heating time is needed to ablate a large area using high intensity focused ultrasound, while multiple-focus pattern or single-focus scan is used. This process may cause unexpected temperature rise in pre-focal regions, which can distort acoustic propagation and affect lesion formations. This article simulates this process by using a thermo-acoustic model, which is based on linear acoustic equations and the heat transfer equation, considering parameter variations due to temperature rising of media on the propagation path. Perfect matched layer (PML) technique is introduced for the absorb boundary condition. Finite difference time-domain (FDTD) method is applied to solve the thermo-acoustic model. The simulations that describe lesion formations in multiple-layer media with different treatment patterns-single focus, multiple-focus and single-focus scan are investigated respectively. These simulations indicate that distortions can be serious with single-focus scan method. Additionally, the single-focus scan patterns applying different interval time length are also simulated. The results imply that the proper interval time can minimize the ultrasound distribution distortion caused by pre-focal heating.
What problem does this paper attempt to address?