Characterization of Golay Detector for the Absolute Power Measurement of Terahertz Radiation

Yingxin Wang,Ziran Zhao,Zhiqiang Chen,Linghui Wang
DOI: https://doi.org/10.1109/irmmw-thz.2012.6380076
2012-01-01
Abstract:The responsivity of a Golay cell detector for continuous-wave 2.52 THz radiation is measured under the modulation frequency from 2 to 50 Hz. Theoretical analysis of the detector output signal is performed by modeling the detection process as a response to a sequence of rectangular pulses. Calibration of the detector with different input powers is achieved by using a thermopile calorimeter as the reference. Our work provides experimental evidence for the power measurement of terahertz radiation with a calibrated Golay detector.
What problem does this paper attempt to address?