Field analysis of benzene, toluene, ethylbenzene and xylene in water by portable gas chromatography-microflame ionization detector combined with headspace solid-phase microextraction.

Jie Ji,Chunhui Deng,Wenwen Shen,Xiangmin Zhang
DOI: https://doi.org/10.1016/j.talanta.2005.11.032
IF: 6.1
2006-01-01
Talanta
Abstract:In this work, portable gas chromatography–microflame ionization detection (portable GC–μFID) coupled to headspace solid-phase microextraction (HS-SPME) was developed for the field analysis of benzene, toluene, ethylbenzene and xylene (BTEX) in water samples. The HS-SPME parameters such as fiber coating, extraction times, stirring rate, the ratio of headspace volume to sample volume, and sodium chloride concentration were studied. A 65μm poly(dimethylsiloxane)-divinylbenzene (PDMS-DVB) SPME fiber, 900rpm, 3.0ml of headspace (1.0ml water sample in 4.0ml vial), and 35% sodium chloride concentration (w/v) were respectively chosen for the best extraction response. An extraction time of 1.0min was enough to extract BTEX in water samples. The relative standard deviation (R.S.D.) for the procedure varied from 5.4% to 8.3%. The method detection limits (MDLs) found were lower than 1.5μg/l, which was enough sensitive to detect the BTEX in water samples. The optimized method was applied to the field analysis of BTEX in wastewater samples. These experiment results show that portable GC–μFID combined with HS-SPME is a rapid, simple and effective tool for field analysis of BTEX in water samples.