Quantitative Analysis Method of BTEX in the Air with Portable GC/MS

Nan ZHANG,Yan ZHENG,Zi-yong WEI,Shu-wei WANG,Hong-li LI,Yan WANG
DOI: https://doi.org/10.3969/j.issn.1003-6504.2012.04.027
IF: 11.4
2012-01-01
Environmental Science & Technology
Abstract:By using the air sampling bag and the preparation of standard gas series,a rapid and simple method for the quantitative analysis of volatile organic compounds in the air with portable GC/MS instrument was determined.In order to test this method,the standard samples of BTEX were analyzed contrastively in different scan modes(full scan and SIM).The results of the comparable experiments showed between full scan mode and SIM mode,the RSD ranges were 5.1%~28.4% and 13%~19%,the recovery ranges were 46%~186% and 60%~110%,the detection limit ranges were 2.504~10.63 μg/m3 and 0.494~2.399 μg/m3 respectively.During the determining process of the trace level contaminants in the air,better results on accuracy and derivation were showed by SIM mode.These conclusions were also imprinted by the determination results of indoor and outdoor samples.
What problem does this paper attempt to address?