Fast determination of ammonia nitrogen and nitrite-nitrogen in tannery effluent by flow injection analysis of auto-conversion

JIANG He-mei,YANG Dong-jing,CHEN Shu-juan,ZHANG Xin-shen
2013-01-01
Abstract:A three-way valve connected before peristaltic pump to modify the flow path of flow injection(FIA) analyzer,it achieved fast and automatic determination of ammonia and nitrite nitrogen with auto-Conversion by three-way valve.With chromogenic system of sulfa-NED,The absorbance was measured at 530nm.the linear range is 8-200μg·L-1 for ammonia nitrogen and 2-50μg·L-1 for nitrite nitrogen;the detection limit for NH4+-N and NO2--N are 0.21μg·L-1 and 0.15μg·L-1,and RSD are 1.48%(n=10)and 1.00%(n=11),respectively.This method has been used to determine NH4+-N and NO2--N in tannery effluent,result shows the recovery is between 98.6%and 101.0%.