Sensitive Wavelength-Modulated Photoacoustic Spectroscopy with DFB Diode Laser

李劲松,高晓明,吴涛,谈图,方黎,张为俊
DOI: https://doi.org/10.3969/j.issn.1000-0364.2007.01.034
2007-01-01
Journal of Atomic and Molecular Physics
Abstract:A new approach to wavelengthmodulation resonant photoacoustic spectroscopy is reported with 2nd harmonic detection, which incorporates room-temperature single mode distributed feedback (DFB) diode laser in the near-infrared. We demonstrate the technique with acetylene detection, yielding a sensitivity limit is about 10 ppm v of acetylene at atmosp heric pressure with 3ms time constant and 3 mW average optical power, which corresponds to the minimum detectable absorption coefficient normalized by laser power and bandwidth is 4.0×10-8 Wcm-1/Hz. Moreover, the optimal pressure condition of buffer gas (N2) has been found to be near 2.4×104 Pa for our photoacoustic system. The present system can effectively eliminate the background signal generated by absorption of the cell windows and by the walls of the sample chamber whose spatial and temporal character are clearly distinct from the resonant wave. We believe that the resonant PA system based on diode laser described here offers substantial advantages over other methods and appears promising as a basis for developing convenient, robust, portable instruments for environmental monitoring and process control.
What problem does this paper attempt to address?