Wear of polyethylene lining and histological profiles of periprosthetic tissue in total hip arthroplasty

WU Wen,YAN Meng-ning,ZHU Zhen-an,DAI Ke-rong
DOI: https://doi.org/10.3969/j.issn.1674-134X.2009.06.013
2009-01-01
Abstract:Objective To investigate the mode of wear through observation on surface of polyethylene(PE) lining by scanning electron microscope(SEM) and to explore the underlying rule between components wear and histological findings of periprosthetic tissue.Methods 12 domestic-made PE linings were retrieved from revision surgery and wear status of the lining surface were detected by SEM.The volume of PE wear was measured with a fluid-displacement method and periprosthetic tissues were collected to make regular H.E.staining and were evaluated with quantified scores.Correlation analysis between PE wear and histological scores were carried out.Results Difference features of wear were identified on surface of PE lining.Crease,ledging and pits were the characteristic plastic wear were commonly seen when cement third-bodies existed,while furrow and scratch were commonly seen when metal third-bodies existed.The volume of PE lining wear correlated well with the histological scores on chronic inflammation of periprosthetic tissues.Conclusions Interposed third-body particles make the PE lining wear more severe and present different features,and wear debris could provoke intense local inflammatory reaction.