Analysis of Finite Element Modeling on Elasto-Plasticity of Polyethylene Liner under Contact Stress after THA

严世贵,吴浩波,陈维善,何荣新,戴雪松,余世策
DOI: https://doi.org/10.3760/j.issn:0253-2352.2004.04.005
2004-01-01
Abstract:Objective To determine the variation in contact stress on polyethylene liner of met-al-backed acetabular components in cases of different cup orientation and femoral head prosthesis diameter un-der different gait states including heelstriking, midstance, toe off and upstairs. Methods Three-dimen-sional finite element models of metal-backed acetabular components and corresponding ultra-high molecular weight polyethylene liners with femoral head diameters of 22 to 32 mm were developed. With the various stress loaded through the head center, contact force of articular surface was analyzed by nonlinear contact elasto-plastic finite element model. Results The cup orientation by different anteversion and/or abduction angles with different head diameters under various loading states affected the resulting contact stress markedly. Liner stress level increased with higher loading states like upstairs and/or smaller femoral head diameter. The peak liner contact stress values were found beyond the yield stress of polyethylene in femoral head smaller than 28 mm under upstairs loading. During walking at same speed on flat ground, the differ-ence of anteversion (0°-30°) and abduction (30°-60°) raised contact stress level of polyethylene liner fluctu-ating within 20% approximately, and all of the peak stress values were less than the yield stress of polyethy-lene. But under upstairs loading state, liner contact stress was found to increase either anteversion or abduc-tion angles. When anteversion angle exceed 15° with various abduction angles (30°-60°), the peak contact stress was all beyond yield stress of polyethylene. Conclusion Among the various designs and operative procedures available for surgeon to select, larger head diameter and proper orientation should be considered for the sake of efficient decrease of the liner contact stress. For patients after total hip arthroplasty, it is nec-essary to avoid or at least reduce higher loading activities.
What problem does this paper attempt to address?