Determination of Diphenylamine in Single-base Propellant by High-performance Liquid Chromatography

LIANG Xiao-dong,CHEN Zhi-yong,QIAO Jing-lin,SUN Li,LI Lin,LUO Chuan-huan
DOI: https://doi.org/10.3969/j.issn.1671-1815.2010.05.024
2010-01-01
Abstract:To establish high-performance liquid chromatography(HPLC)method determination of content of diphenylamine(DPA)in propellant.Venusil MP-C18 column(250 mm×4.6 mm,5 μm)with methanol-water(80:20)as mobile phase,the flow rate was 1 mL/min,the ultraviolet detection wavelength was 280nm.There was a good linear relationship within(0.04—80)μg/mL,regression equation was Y = 708 956 X,r=0.999 8.The method is validation,and the stability tests under various conditions are performed.The reproducibility and precision of method are good(RSD=0.64%).The lowest detectable concentration of diphenylamine is 0.05 μg/mL.The recoveries are 99%~101% and the relative standard deviation is less than 1%.It is conclusied that the method proposed for content of diphenylamine by HPLC is simple,accurate,specific and could be used in quality control diphenylamine.
What problem does this paper attempt to address?