Direct Determination of Micro Organic Amine in Water by Gas Chromatography

Hou Xiuhua,Gao Minghua,Liang Yun,Zhao Pu
DOI: https://doi.org/10.3969/j.issn.1006-1878.2006.02.021
2006-01-01
Abstract:Some kinds of micro organic amine in water were direct determined using gas chromatograph with hydrogen flame ionization detector(FID).The samples were filtrated and direct injected into the CP7448 capillary column,and the micro content of methylamine,dimethylamine,ethylamine and diethylamine were determined by external standard method.Under the selected concentrations,there is a good linear relationship with peak area within a range of mass concentration.The recoveries of the 4 components are 96.7%,99.0%,101.2% and 99.2%;The relative standard deviations of the method are(4.3%,)(3.7%,)(5.5%) and(3.1%;)The minimum detection mass concentrations are 2.0,2.0,1.5 and 1.5 mg/L,respectively.This method is simple with high recovery and precision.
What problem does this paper attempt to address?