A New Electroplated Ir/Ir(OH) X Ph Electrode and Its Application in the Coastal Areas of Newport Harbor, California
Xiao Zhang,Ying Ye,Yating Kan,Yuanfeng Huang,Jianjun Jia,Yue Zhao,Chen-Tung Arthur Chen,Huawei Qin
DOI: https://doi.org/10.1007/s13131-017-1064-5
2017-01-01
Acta Oceanologica Sinica
Abstract:Resulting from the rising levels of atmospheric carbon, ocean acidification has become a global problem. It has significant impacts on the development, survival, growth and physiology of marine organisms. Therefore, a highprecision sensor is urgently needed to measure the pH of sea-water. Iridium wire with a diameter of 0.25 mm is used as the substrate, and an Ir/Ir(OH) x pH electrode is prepared by a one-step electrochemical method in a Li OH solution at the room temperature. A scanning electron microscope(SEM) observation reveals that it is coated with nanoscale particles. In laboratory tests, the electrode exhibits a very promising pH response, with an ideal Nernst slope(56.14–59.52), fast response, good stability and long life-span in tested pH buffer solutions. For a sea trial,four pH electrodes and one Ag/Ag Cl reference electrode are integrated with a self-made chemical sensor, and a profile detection of nearly 70 m is implemented near Newport Harbor, California on August 3, 2015. The results reflect that the pH value measured by the sensor is very close to the data given by Sea-Bird 911 plus CTD, with a difference value ranging from 0.000 075 to 0.064 719. And the sensor shows a better data matching degree in 0–40m water depth. In addition, the high precision and accuracy of the sensor make it possible to use in the ocean observation field.