The Effect of Reconstruction Parameters of Spine QCT on Bone Mineral Density

WANG Yusheng,GUO Zhe,ZHAO Haizhu,WANG Ce,ZHANG Lei,ZHAO Yingwei,YU Aihong,CHENG Xiaoguang
DOI: https://doi.org/10.3969/j.issn.1005-5185.2011.12.004
2011-01-01
Abstract:Purpose To optimize the scan protocol for accurate quantitative computed tomography(QCT) bone mineral density(BMD) measurements with low radiation dose.Material and Methods European Spine Phantom(ESP) and QCT calibration Phantom(Image Analysis,Inc,US,G2208) were scanned with a Toshiba 16-sclice CT scanner.Tube voltage and tube current were 120KV and 50mA to 350mA.Slice thickness was 1 x 16mm with table height of 73cm.Scan range was 10cm.FOV was 40cm.Reconstruction matrix was 512×512 and reconstruction interval was 0.8cm.Dose length product(DLP) was recorded.Images of ESP were reconstructed using standard algorithm(FC 03) and bone algorithm(FC 30) .Reconstructed data were transferred to BMD workstation.BMD was measured using QCT PRO TM 3D spine BMD application module(version 4.0,mindways software Inc) .Statistical analysis was performed using SPSS 15.0.Results The correlation coefficient of BMD with mAs of FC03 and FC 30 were 0.401(P<0.05) and 0.216(P>0.05) ,respectively.The correlation coefficient of mAs with DLP is 0.999(P<0.01) .The difference of BMD between FC03 and FC30 was statistically significant(t=-13.089,P<0.01) .There were statistically significant differences between the BMD of FC 03 and standard value of ESP(t=-19.220,P<0.01) .The BMD of FC 30 and standard value of ESP were not statistically different(t=0.506,P=0.616) .Conclusion Reconstruction parameters of QCT affect the measured value of BMD.Images reconstructed using bone algorithm(FC 30) exclude interfering factors and increase the accuracy of the BMD measurement.Reconstruction parameters should be taken into account when the scan protocols were formulated.
What problem does this paper attempt to address?