Determination of Titanium in Sea-Water Using Adsorptive Cathodic Stripping Voltammetry

H LI,CMG VAN DEN BERG
DOI: https://doi.org/10.1016/s0003-2670(00)81963-7
IF: 6.911
1989-01-01
Analytica Chimica Acta
Abstract:A procedure is presented for determining titanium dissolved in sea water and fresh water using cathodic stripping voltammetry preceded by adsorptive collection of its mandelic acid complexes. The optimized conditions include a pH of 3.3, a mandelic acid concentration of 4 mM and an adsorption potential of −0.15 V. The peak potential is −0.9 V. Differential pulse modulation is used, with a pulse rate of 10 s−1 and a scan rate of 20 mV s−1. The limit of detection is 0.03 nM Ti when a 10-min adsorption time is used. The sensitivity is higher for fresh water than sea water owing to interfering Ti complexation by fluoride in sea water. Interference caused by surface-active organic compounds, which lower the sensitivity owing to adsorption on the mercury electrode, is removed by UV photolysis of the sample prior to the analysis. The method was tested by determining Ti in samples from the North Sea and the Indian Ocean.