Abstract:Optimum conditions of determining trace tin by anodic stripping voltammetry were investigated by using a Nafion/MWCNT modified glassy carbon electrode.The result indicated a sensitive oxidation peak was observed and the peak potential of anode was-570 mV in NH4Cl-HCl electrolyte with pH being 1.2,provided accumulation time was 3 min and potential scan rate was 200 mV/s.The trace tin was determined by using this peak.There was a good liner relationship between the peak current and Sn(Ⅳ) concentration in the range of 4.0×10-8 ~1.0×10-6 mol/L with correlation coefficient of 0.997,and the detection limit was 1.0 × 10-8 mol/L.This propose method had been applied to the determination of actual water sample and the average recovery was 97.48%.The result was satisfying.