Trace Cu2+ detection based on GH-PEDOT:PSS-Pt NP-modified glassy carbon electrode
Hao Changshi,Wang Yiding,Wu Hongpeng,Duan Shaojing,Liu Bo,Yan Luting
DOI: https://doi.org/10.1007/s10854-023-11882-w
2024-01-05
Journal of Materials Science Materials in Electronics
Abstract:We successfully developed a GH-PP-Pt/GCE electrode by compounding graphene hydrogel (GH) with poly(3,4-ethylenedioxythiophene)-poly(styrenesulfonate) (PEDOT:PSS), and in situ electrodeposition of Pt nanoparticles. The experimental results prove that the performance of GH-PP-Pt/GCE to trace copper ions significantly improved, the sensitivity reaches 45.673 μA/μmol L −1 , the limit of detection is as low as 9.9 nmol L −1 , and the linear range is 0.08–10 μmol L −1 . Its detection ability is at a relatively high level among the existing Cu 2+ electrochemical detection sensors, and the repeatability, stability, and anti-interference are good.
engineering, electrical & electronic,materials science, multidisciplinary,physics, condensed matter, applied