Determination of Trace Cadmium with Nafion/bismuth Modified Screen-Printed Electrode

唐立超,唐美华,石成成,蔡清,张东明,张之翼,王济奎,陈国松
DOI: https://doi.org/10.16606/j.cnki.issn0253-4320.2014.08.009
2014-01-01
Abstract:Nafion modified screen-printed electrode in combination with in-situ bismuth film is fabricated to determine trace cadmium. Nafion is dropped onto the surface of a screen-printed electrode and dispersed to form a modified layer. Then Bi3 +and Cd2 +are electrodeposited together on the surface of the modified electrode in 0. 1 mol /L HAc /NaAc solution( pH 4. 7). The concentration of cadmium is measured by square wave stripping voltammeter( SWSV). The effects of Nafion and concentration of bismuth on the stripping peak current are investigated. Under the optimal condition,there is a good linear relationship between the concentration of cadmium and the stripping peak current in the range from 0 to 100 μg /L of cadmium. The detection limit is 1 μg /L. This method has good accuracy,sensitivity and reproducibility. It can be applied in the rapid determination of trace heavy metal ions in rice samples.
What problem does this paper attempt to address?