Voltammetric Determination of Trace Palladium by Carboxylic Carbon Nanotubes Modified Glass Carbon Electrode

LI Jun-hua,KUANG Dai-zhi,FENG Yong-lan,TANG Wen-qing,QU Jing-nian
DOI: https://doi.org/10.3969/j.issn.1000-7571.2010.12.002
2010-01-01
Abstract:A new mercury-free determination method of trace palladium was established based on carboxylic multiwall carbon nanotubes(c-MWCNT) modified glass carbon electrode.In 0.1 mol/L HAc-NaAc(pH 4.5) buffer solution,there was a sensitive anodic stripping peak on linear sweep voltammograms and the stripping peak potential was 0.162 V when accumulation time of Pd2+ at the modified electrode was 30 s at-1.1 V and potential scan rate was 100 mV/s.There was a good linear relationship between peak current and Pd2+ concentration in the range of 5.0×10-10-1.0×10-7 mol/L with correlation coefficient of 0.999.Detection limit of this method was 1.5 × 10-10 mol/L.This kind of modified electrode could be used in the determination of palladium in palladium catalyst with recoveries of 98%-103% because of its good stability.
What problem does this paper attempt to address?