Determination of Cr and Ni in Seawater by High-Resolution Continuum Source Electrothermal Atomic Absorption Spectrometry

M. Yu. Burylin,E. S. Kopeyko,E. S. Kostyuchenko
DOI: https://doi.org/10.1134/s1061934824030043
IF: 1.237
2024-04-03
Journal of Analytical Chemistry
Abstract:Conditions for the determination of Ni and Cr in seawater by high-resolution continuum source electrothermal atomic absorption spectrometry ( HR-CS ETAAS ) are developed. Nitric acid is used as a chemical modifier. The analytical capabilities of this method are compared with those of the version of low-resolution line source electrothermal atomic absorption spectrometry ( LR-LS ETAAS ) with deuterium background correction. A calibration dependence is constructed on standard analyte solutions. Procedures for determining Cr and Ni are tested on water samples from the Black Sea using the spiked test. Using the HR-CS ETAAS method, limits of detection (μg/L) of 0.12 for Cr and 0.09 for Ni (dosage 80 μL) are achieved. For the LR-LS ETAAS method, these indicators are worse: 0.6 (Cr) and 1 μg/L (Ni) with a dosed volume of seawater sample of 10 μL. A possibility of using the ETAAS method for the direct determination of Cr and Ni contents in seawater at a level two orders of magnitude lower than their maximum permissible concentrations is demonstrated.integral absorbance
chemistry, analytical
What problem does this paper attempt to address?