Determination of ultra-trace level Cr(VI) in seawater using large-volume direct injection by LC-ICP-MS with in-line matrix elimination

Zhenzhen Yao,Bingru Li,Zhihong Ma,Beihong Wang
DOI: https://doi.org/10.1039/d3ja00191a
2024-02-14
Journal of Analytical Atomic Spectrometry
Abstract:In this study, we aimed to improve the detection of hexavalent chromium (Cr( VI )) in seawater by using large volume injection and in-line matrix elimination device in conjunction with LC-ICP-MS. Firstly, the sample solution was introduced into the chromatographic separation system through a 1000 μL quantitative loop. Then, the chromatographically separated solution was precisely introduced into the ICP-MS for analysis at the time of the Cr( VI ) peak (2.8–4.2 min) by the valve switch in the in-line matrix elimination. And the separation of Cr( VI ) was performed on a Dionex IonPac AG7 analytical column (50 mm × 4 mm, 10 μm) with 60 mmol L −1 NH 4 NO 3 as the mobile phase at a flow rate of 1.0 mL min −1 for 5 min. The detection limits (LOD) and limit of quantitation (LOQ) were found to be 0.002 and 0.007 μg L −1 , respectively. Method precision and accuracy were satisfactory at 3.5% relative standard deviations (RSDs) for 0.1 μg L −1 of Cr( VI ), and 90% ∼ 105% recoveries for real samples. The developed method was successfully applied for the analysis of Cr( VI ) in environmental seawater samples.
spectroscopy,chemistry, analytical