Simultaneous Determination of Trace Cr(Iii) and Cr(Vi) in Water Using Ion-Pairing Reversed Phase Chromatography-Inductively Coupled Plasma Mass Spectrometry

Wang Hua-Jian,Li Yan-Hong,Feng Wei-Yue,Wang Meng,Jia Guang,Wang Bing,Zhu Mo-Tao,Wang Yun,Chai Zhi-Fang
DOI: https://doi.org/10.3321/j.issn:0253-3820.2009.03.025
2009-01-01
Abstract:A method was developed to simultaneously determine trace chromium species in water using ion-pairing reversed phase chromatography (RPIPC)-inductively coupled plasma mass spectrometer (ICP-MS). The chromatographic parameters were optimized to get the best separation, including the pH value of mobile phase, the concentrations of ion-pair reagent and methanol and the addition of EDTA in the mobile phase. The optimium mobile phase was 2.0 mmol/L tetrabutlyammonium (TBA), 5% methanol (v/v), pH = 5.5. Using collision cell technology to eliminate the mass interferences of (ArC+)-Ar-40-C-12 and (ClOH+)-Cl-35-O-16 in the analysis of Cr-52(+), and the detection limits of Cr(III) and Cr(VI) were 0.15 mu g/L and 0.16 mu g/L for 100 mu L injections. The recovery range of spiked Cr(III) and Cr(VI) in water samples was 93.6% - 106.2% and the RSDs were less than 4% (n=3). The results were satisfying when analyzing the chromium speciation in drinking water and rain water using this method.
What problem does this paper attempt to address?