Test Running and Data Analysis on the Dual Mixer Time Difference System at NTSC

WANG Zheng-Ming,DONG Shao-Wu,LIU Chun-Xia,QI Yi
2006-01-01
Abstract:The newly imported dual mixer time difference system(DMTD) manufactured by Timetech Company has been tested.The operating theory and the structure of the device are described.The frequency signal of the master clock(MC) at NTSC Standard Time and Frequency Laboratory is used as the reference and the frequency signals output of 5 hydrogen masers and 18 cesium frequency standards are fed to the ports of the DMTD for phase comparison with MC.The signals out of a frequency distribution amplifier the input of which is the frequency signal of MC are used to test the measuring precision and frequency stability of the DMTD.The comparison between using DMTD and using the time interval counter,SR 620,is given.The measured time difference fluctuations between MC and each of the clocks show that it is better to use such kind of measuring system for the time difference measurement between hydrogen masers.The DMTD will be used for time keeping at NTSC time lab and for the frequency comparison between hydrogen masers and a cesium fountain in the near future as well.
What problem does this paper attempt to address?