Safety experiment for loss of vacuum on a LHe vessel

Shen Biao,Yu Lihong,Zhang Peng,Wang Ruzhu
DOI: https://doi.org/10.3969/j.issn.1001-7100.2007.03.003
2007-01-01
Abstract:Sudden loss of the vacuum of a liquid helium container raises serious safety concerns.Following the gas venting,dramatic heat leakage into the LHe is expected to cause a severe pressure rise due to the boiling of the helium.In addition,cold helium vapor of extremely low temperatures could result in catastrophic damage if not appropriately disposed.In this paper,we present the results of helium and air exposure experiments on a LHe vessel of a volume of approximately 10 liters with various insulation designs,namely,absence of insulation,20 layers or 50 layers of super-insulation.It is found that,the installation of super-insulation could work as a thermal buffer between the room temperature gas entering the vacuum space and the LHe after the insulating vacuum is compromised,effectively reducing the heat inflow into the cryogenic vessel.Also,the results show that,as far as the LHe is concerned,venting air into the insulating vacuum brings about less serious damage than venting helium.