A Real-time Synchrotron Radiation CT System for Materials Analysis in Tension/Compression Testing

Xia Yuan-ming
2006-01-01
Abstract:A real-time synchrotron radiation CT system for materials analysis in the tension/compression testing was introduced.By using this system,the macro P-Δ curve of the specimen can be obtained,as well as the micro CT photograph at any loading state.One of the key components of this system was the integral loading/rotating device.Based on the pixel size(10.9μm×10.9μm) of the X-ray CCD, The kinematic positioning accuracy requirement was given through analysis.In the device,a four-post frame was adopted,two synchronously rotating grip holders were used to realize rigidity rotation of the specimen,and the axial movement of the upper grip holder implemented the tension/compression loading.Positioning accuracy in the axis,radial and circumferential directions were achieved by adopting many advanced and delicate technologies.Measurement on the device indicated that the actual positioning accuracy of the device was better than the kinematic positioning accuracy.Evaluation test on the system indicated that the spatial resolution is better than the pixel size without considering the noise.