SINGLE STANDARD CALIBRATION METHOD BY GC-ECD SYSTEM ON THE BACKGROUND N_2O MEASUREMENT

方双喜,周凌晞,汪巍,张芳,姚波,许林,刘立新,温民
IF: 4.236
2011-01-01
Environmental Chemistry
Abstract:In this study,non-linearity correction based on linear regression was applied to the N2O measurement in air sample.Three ambient N2O samples were calibrated several times against 278.11×10-9 standard by this method.Average difference to that of the polynomial fitting results was less than 0.1 ×10-9 and precision was better than(0.03±0.09)%,proving proved that the method could roughly correct error from the linear regression.Additionally,the method could reduce the number of working standards and their frequency of injection.It can also obviously increase measurement efficiency of ambient air.