Determination of High Concentration of Ca(Ⅱ) in Geothermal Water with Flow Injection-Micro-flow-through Ion Selective Electrode

WANG Zhi-cheng,LI Yong-sheng
2012-01-01
Abstract:The FI-micro-flow-through Ca(Ⅱ) selective electrode(FI-MF-Ca(Ⅱ) ISE) was applied to the determination of high concentration of Ca(Ⅱ) in geothermal water.Various influential factors were studied.The optimized conditions were as follows: ① length of the mixing coil: 30 cm;② flow rate of the carrier: 0.99 mL·min-1;③ sampling volume: 60 μL;④ total ionic strength adjusting buffer solution(pH 9.7): a mixed solution containing 0.02 mol·L-1 sodium borate,0.02 mol·L-1 sodium hydroxide,10 mg·L-1 Ca(Ⅱ) and 0.60 mol·L-1 sodium chloride,with a flow-rate of 0.88 mL·min-1.Linearity range of the method was kept between 0.16-20.00 g·L-1.Precision of the method was tested at the concentration levels of 1.00 and 2.00 g·L-1 of Ca(Ⅱ) for 11 determination.Values of RSD′s were found to be 2.8% and 1.8%,respectively.The proposed method was used in the analysis of samples of geothermal water,giving values of recovery by standard addition method in the range of 102%-111%.
What problem does this paper attempt to address?