The Development of a Flow Injection Analysis System Based on Iodide Ion Selective Electrode

LI Nan-qi,ZHANG Ping,CHEN Song-zhe,WANG Lai-jun,XU Jing-ming
DOI: https://doi.org/10.3969/j.issn.1000-0720.2011.06.012
2011-01-01
Abstract:The Iodine-Sulphur thermo-chemical water-splitting process is considered as one of the most promising way of massive hydrogen production.For monitoring and controlling the process conditions or carrying out relative kinetics study,it is essential to determine the concentration of I-online.A flow injection analysis system with I-ionic selective electrode(ISE) as a detector was designed and constructed.The main factors influencing the electrode performance,such as temperature and ionic strength were studied,and the operational parameters including carrier flow rate and injection volume were optimized.The performance of the device was evaluated,the results show that the detection limit for iodine ion is about 2mg/L,and the sampling rate reaches at least 40/h,or 60/h for lower concentration samples.Therefore,the device could meet the demand for online determination of I-.In addition,the device may be used to online determine the concentration of H+ in the range of 0.1-2mol/L.
What problem does this paper attempt to address?