Application of Portable Electrokinetic Flow Analysis System for Determination of Chromium(Vi) and Cadmium (ii) in Mineral Water

L Yang,YZ He,W Gan,N Deng,M Li,XQ Lin
DOI: https://doi.org/10.3321/j.issn:0253-3820.2001.05.015
2001-01-01
Abstract:The improvement of the electroosmotic;pump properties and the effects of the bi-directional electrostacking system on the pre-concentration factor were investigated. A portable electrokinetic flow analysis system with electroosmotic driving and bi-directional electrostacking unit was introduced. The improved system with graphite furnace atomic absorption spectrometry was employed to separate, pre-concentrate and determine Cr(VI) and Cd(II) in mineral water. The detection limit of Cr(VI) and Cd(II) was 9 ng/L and 10 ng/L (3 sigma of blank, n = 11), respectively. Their recoveries of 200 ng/L Cr(VI) and 100 ng/L Cd(II) added to samples were (105 similar to 107) +/- 2% and (104 similar to 106) +/- 2%( n = 3).
What problem does this paper attempt to address?