Determination of total inorganic carbonate in water samples by a novel gas diffusion unit coupled with flow injection analysis

Yanlin Wei,Jinming Lin,Motomizu Shoji
2005-01-01
Abstract:A novel gas diffusion unit and the preparation method of less carbonate water were developed. The gas diffusion unit was constructed with double tubing: the inner tubing was a micro porous PTFE (polytetrafluoroethylene) tubing (1.0 mm inner diameter and 1.8 mm outer diameter, pore size 2 μm, porosity 50%); the outer tubing was made of glass with 2.0 mm inner diameter. It was coupled with flow injection analysis (FIA) and used for the determination of total inorganic carbonate (TIC) with satisfactory results in various water samples, such as purified water, rain water, tap water, river water, sea water and alkali water solutions. The calibration graphs were linear over a wide dynamic range from 10 -6 M to 10 -3 M. The limit of detection (LOD) corresponding to a signal-to-noise ratio of three was 1×10 -6M of total inorganic carbonate.