Improvement in Exhaust Method of Exadrop Precision Infusion Set

Sun Qing-zhu
2011-01-01
Abstract:Objective To improve the exhaust method for Exadrop precise infusion set so as to reduce the bubble generation rate.Methods From July to October 2010,50 respiratory patients with Exadrop precision infusion set were divided into traditional group and improvement group.Following the improved and traditional exhaust methods,the exhaust effects were compared between the two groups.Results In the improvement group,there were 94 times of total exhaust and the bubble rate after exhaust and the bubble rate during infusion were 12.8%(12/94) and 12.8%(12/94) respectively.In the traditional group,there were 92 times of total exhaust and the bubble rate after exhaust and the bubble rate during infusion were 67.4%(62/92) and 54.3%(50/92) respectively.The bubble rate after exhaust(χ2=24.22) and the bubble rate during infusion(χ2=18.09) in the improvement group were significantly higher than those of the traditional group(P0.01).Conclusion The improved exhaust method of precision infusion set can effectively eliminate residual bubbles,avoid the increased nursing workload due to repeated adjustments,reduce the residual gas during infusion,ensure the safety of patients and increase the patients' satisfaction.