Design and application of a new type of medical drainage device in the measurement of ICU intra-abdominal pressure

董莲莲,陈香萍,劳月文,张奕
DOI: https://doi.org/10.3760/cma.j.cn211501-20190902-02486
2020-01-01
Abstract:Objective:To explore the clinical application of a new type of medical drainage device designed to measure the intra-abdominal pressure of ICU patients.Methods:The 65 patients with severe acute pancreatitis treated in our hospital from April to September 2018 were selected as the experimental group; the 54 patients with severe acute pancreatitis treated from October 2017 to March 2018 were selected as the control group. Patients in the control group used traditional drainage bags to measure intra-abdominal pressure, while patients in the experimental group used a new-designed medical drainage device to measure intra-abdominal pressure. Compare the cost of consumables used for the first time in the experimental group with the control group, the incidence of acupuncture injuries, the incidence of urethral leakage, the incidence of rupture of the catheter balloon, and the satisfaction of the nurses. The 32 patients admitted from January to March 2018 in the control group were measured again using a new drainage device after the abdominal pressure measurement operation was stopped in order to compare the accuracy of the two methods.Results:The per capita consumable cost of the experimental group was 5.71 yuan, 0 cases of needle stick injury, 0 case of catheter leakage, which were lower than the control group (22.36 yuan, 1 case, 7 cases), the difference was significant ( P<0.05). The nurse operation satisfaction score was 13.85±0.93, which was higher than the control group (10.00±1.05). The difference was statistically significant ( t value was -20.323, P<0.05). Conclusion:In the operation of intra-abdominal pressure measurement, the use of a new type of medical drainage device can ensure the accuracy of the measurement, reduce the cost of consumables, needle stick injuries and the incidence of urinary catheter leakage, and improve nurse operation satisfaction.
What problem does this paper attempt to address?