Change of the thresholds of helix electrodes during the implantation

盛琴慧,杨俊娟,杨平,丁燕生
2008-01-01
Abstract:Objective To investigate the threshold variation of the helix electrodes during the implantation. Methods Seventy-two patients with 74 helix electrodes were enrolled in our study.After the cardiac pacing leads were put on the suitable sites of the ventricle or the atrium,the thresholds were tested when the electrode leads were screwed in before and after 0,5,and 10 minutes. Results The thresholds of the helix electrodes were: 0.56±0.30 V before screwed,0.97±0.49 V,0.46±0.29 V and 0.36±0.20 V after screwed immediately,5 and 10 minutes after screwed later. Conclusion The thresholds of helix electrodes increased significantly after being screwed immediately,but decreased gradually to the original level before being screwed after 5-10 minutes.