Relationship between Using Frequency and Transesophageal Cardiac Electrophysiology Pacing Threshold Voltage of E-sophageal Electrode Catheter

Wei CHU,Cui-zhen YUAN,Yuan XU
DOI: https://doi.org/10.3969/j.issn.1007-9572.2014.32.025
2014-01-01
Chinese General Practice
Abstract:Objective To explore the relationship between the frequency of use of esophageal electrode catheter and transesophageal cardiac electrophysiology pacing threshold voltage. Methods A total of 123 patients who had esophageal pacing check during 2012 - 2013 in Daping Hospital were enrolled in this study. This study compared the pacing thresholds in groups A (having used for 1 - 2 times),B(having used for 3 - 4 times),C(having used for more than 4 times). Results The aver-age pacing threshold voltage was(13. 07 ± 2. 63)V in group A,(17. 15 ± 2. 60)V in group B,(21. 65 ± 1. 89)V in group C. The difference was significant(F = 147. 019,P < 0. 05);and group B was higher than group A,group C was higher than group A,B(P < 0. 05). Conclusion Esophageal electrode catheter will lead to the increase of pacing threshold due to passi-vation coating of surface of the electrode when it using for more than 3 - 4 times,which hinders its promotion and use in clinical practice.
What problem does this paper attempt to address?