Study on the relationship between amplitude of intrinsic R waves and evoked responses measured by pacemakers with an autocapture pacing system

YUAN Cui-zhen,XU Yuan,GUO Ji-hong,ZHAO Jing
DOI: https://doi.org/10.13333/j.cnki.cjcpe.2007.02.021
2007-01-01
Abstract:Objective To evaluate the relationship between evoked responses and amplitude of intrinsic R waves measured by pacemakers with an autocapture pacing system. Methods 161 consecutive patients who had undergone implantation of pacemakers with an autocapture pacing system were included in the study. The amplitude of intrinsic R waves (mV), impedance, threshold and evoked responses (mV) were measured by standard pacing system analyzer, then analyzed the relativity between these factors and ER. Results The amplitude of R waves (13.08±4.82 mV) and 99%CI of evoked responses (11.90±15.01 mV) were measured intraoperatively. Correlation coefficient (r) was 0.05 (P>0.05). Conclusion There was no association between amplitude of intrinsic R waves and evoked responses. ER shoud be measured separately for safety.