Clinical assessment of pacemaker replacement

CAO Wei-e,CUI Ke-jian,ZHENG Hong-chao,WANG Wei,PAN Cheng,CAO Jia-qi
2008-01-01
Abstract:Objective To investigate causes for pacemaker replacement,life span of the electrodes and operative complications.Methods Pacemaker replacement occurred in 90 patients including 42 males and 48 females with a mean age of 75.69±9.17 years,including four patients between 47 and 59 years.Fifty patients had sick sinus syndrome and 40 patients had third degree atrioventricular block.Replacement was performed by VVI mode in 77 patients,by DDD mode in 7 patients,and by AAI mod in 5 patients.Replacement was transferred from VVI to DDD mode in one patient.Seventy-eight patients underwent replacement for the first time,8 patients for the second time,and 4 patients for the third time.Results The life span of the pacemakers in these patients was(110.17±30.25)months,and that of the electrodes was(134.32±41.86)months.The battery was normally exhausted in 75 patients and ahead of schedule in 14 patients.In one patient,the electrodes were replaced while the original pacemaker was stilled in use.The threshold during pacemaker replacement was significantly higher than initial pacing threshold(0.97±0.52)V vs(0.76±0.36)V(P<0.01).The impedances of the electrodes during pacemaker replacement were lower than initial impedances(687.64±153.13)Ω vs(750.56±252.17)Ω(P<0.05).There was dislocation of electrodes in one case and hematoma of the pacemaker sack in one case and the incision cure delayed in two cases.Conclusions The primary cause for pacemaker replacement was battery exhaustion.In most cases,the original electrodes were still usable when the pacemaker was replaced for the first time.