Study on the Influence of a Micro Axis Flow Blood Pump Diffuser on Flow Field

CHEN Jianzhong,ZHANG Xiwen,HE Feng,BAI Jing
DOI: https://doi.org/10.3969/j.issn.1002-3208.2007.01.001
2007-01-01
Abstract:Based on computational fluid dynamics(CFD) technology, laser Doppler velocimetry and high speed CCD flow field measuring methods, the flow fields of a micro axis flow blood pump under several typical working conditions were studied. The details of the flow fields such as the backflow, vortex and secondary flow were revealed. After comparing the flow fields between diffuser A and B in the experiment, the flow field of diffuser A was found to be planar in evidence, which met the blood pump flow field’s need better than that of diffuser B. This work is valuable to optimize a blood pump diffuser.
What problem does this paper attempt to address?