Influence of Diffuser for Artificial Axial Heart Pump

Ying-Peng WANG,Xin-Wei SONG,Chun-Tong YING
DOI: https://doi.org/10.3969/j.issn.0258-8021.2007.05.021
2007-01-01
Abstract:Inducer and diffuser play important roles when blood is transported by an artificial axial heart pump.In this article,computational fluid dynamics(CFD) simulations were applied to analysis the influence of diffuser for the axial artificial heart pump.The current work was focused on the change of static pressure rise,efficiency,working situation and fluid shear stress for pump.Analysis results indicated that proper diffuser model could improve the fluid distribution along the pump,project the direction of outlet velocity parallel to axis,keep good pump efficiency and supply enough static pressure,which are necessary for blood transportation.In conclusion,it is very important to add a diffuser to the impeller,while strict selection for parameters of diffuser must be carried out during the whole design procedure of axial artificial heart pump.
What problem does this paper attempt to address?