DETERMINATION OF RELEVANT SUBSTANCES OF CEFACLOR SUSPENSION BY HPLC

Xiang Shiying,Lin Bo,Liang Zhenyi
DOI: https://doi.org/10.3969/j.issn.1008-6145.2009.01.013
2009-01-01
Abstract:A HPLC method for the determination in relevant substances of cefaclor suspension was established with Hypersil C18 BDS column.The mobile phase A was 0.78% sodium dihydrogen phosphate solution(pH 4.0).The mobile phase B was 0.78% sodium dihydrogen phosphate solution-acetonitrile(volume ratio was 55:45),with flow rate of 1 mL/min.The detection wave lenth was at 220 nm.The relative standard deviation of detection results was 0.01%(n=6).The relative error of marker specimen detection results was not more than 1.0%.The method can need the enterprise equirements of quality control testing.
What problem does this paper attempt to address?