Flow Injection Analysis of Cefaclor with Enhanced Chemiluminescent Detection

SF Li,H Cui,YG Sun,XQ Lin
DOI: https://doi.org/10.3969/j.issn.1000-0720.2006.06.003
IF: 1.193
2001-01-01
Chinese Journal of Analytical Chemistry
Abstract:A novel flow injection method was developed for the determination of cefaclor based on the enhancement of cefaclor on the chemiluminescence of Luminol-H_2O_2 system.The method is simple,selective and sensitive with a detection limit of 6.0×10~(-9) g/mL.It is effective to determine cefaclor in the range of 1.0×10~(-8)~2.0×10~(-5) g/mL.The relative standard deviation is 2.2% for the determination of 1.0×10~(-6) g/mL cefaclor(n=11).The proposed method has been successfully applied to determination of cefaclor in Xikelao granules with satisfactory results.
What problem does this paper attempt to address?