Study on Reliability of Resting Metabolic Rate Determined by Fitmate Lung-Function Analytical Instrument

LI Min,PIAO Jian-hua,LIU Jian-min,TIAN Ying,SUN Rui,YANG Xiao-guang
DOI: https://doi.org/10.3969/j.issn.1004-8456.2007.06.009
2007-01-01
Abstract:Objective To assess the validity and reliability of RMR determined by FitMate.Method A validation study of FitMate was devised to compare with the Douglas-Haldane method, which was the classical method for gas metabolism. Sixteen healthy graduates at the age of 23 to 28 years were randomly selected, including eight males and eight females.Results The difference of RMR determined by Douglas-Haldane and FitMate (P0.05) was not found, the P value of paired t test of RMR of male and female were 0.402?7 and 0.635?8 , respectively. By correlation analysis, the correlation coefficients of RMR of male and female determined by two instruments were 0.75(P= 0.003?3 ) and 0.63(P= 0.011?2 ), respectively, which had significant correlation (P0.05).Conclusion The FitMate could make reliable RMR measurement.