Validity of a portable spirometer based on differential pressure sensor

Zou Kun,Ni Chaomin,He Zijun
DOI: https://doi.org/10.3969/j.issn.1000-0399.2012.12.010
2012-01-01
Abstract:Objective To assess the validity of a portable spirometer,compared with a diagnostic spirometer(Medisoft BodyBox S5500,Belgium).Methods A group of 45 patients(mean age 52.8±18.67 years,27 male)were tested with both a portable spirometer and Medisoft BodyBox S5500,according to the guidelines of the American Thoracic Society.Then the results of the FVC,FEV1,FEV1/FVC,PEF,MEF75,MEF50,and MEF25 obtained from both spirometers were contrasted.Results Pulmonary function measurements from the portable spirometer correlated closely with those from the Medisoft BodyBox.The r of those parameters was 0.97,0.977,0.853,0.929,0.929,0.868,and 0.857 respectively for FVC,FEV1,FEV1/FVC,PEF,MEF75,MEF50,and MEF25.Bland-Altman chart showed good limits of agreement between the two apparatus,only two or three data were beyond the limits of agreement.The differences were statistically significant for FVC,FEV1,PEF,MEF75 from both apparatus(P<0.05).Conclusion The portable spirometer can be used for testing lung ventilation function.