Evaluation of a Thermal-Based Flow Meter for Assessment of Mobile Resting Metabolic Rate Measures.

Nai-Yuan Liu,Yue Deng,Francis Tsow,Devon Bridgeman,Xiaojun Xian,Jane J. Dean,Janet L. Wilson,Nongjian Tao,Doina Kulick,Erica Forzani
DOI: https://doi.org/10.1155/2018/9186475
IF: 2.336
2018-01-01
Journal of Sensors
Abstract:This work evaluates the use of a new flow meter to assess exhalation rate. A mobile indirect calorimeter (MIC) was designed and used to measure resting metabolic rate (RMR), which relies on the measure of O-2 consumption rate (VO2) and CO2 production rate (VCO2). The device was produced from a commercially available and well-established indirect calorimeter and implemented with a new flow meter for the purpose of this study. VO2 and VCO2 were assessed by measuring exhalation rates using the new flow meter and O-2 and CO2 concentrations in breath using the original colorimetric sensors of the indirect calorimeter. The new flow meter was based on a thermal flow meter (TFM) affixed to an orifice with a diameter of 6.8 mm used as a passage for exhaled breath from 16 subjects. The results were compared with a metabolic cart (Medical Graphics), which was connected in series to the modified device. We found that 69% of the results had more than a 10% difference between the modified MIC device and the reference instrument, suggesting that the sensitivity of the thermal flow meter changed over time, which precluded its use as a flow meter for breath flow rate measurement.