Analysis of the Medical Linear Accelerator Output Stability

JIANG Rui-yao,XIONG Fei,HUANG Guo-feng,LI Zhao-bin,FU Shen
DOI: https://doi.org/10.3969/j.issn.1005-202x.2011.05.007
2011-01-01
Abstract:Purpose: This paper tries to analyze the output stability and the main influent factors of a medical linear accelerator. Methods: With the help of SPSS15.0,the paper statistic the output dose of 6 MV、15 MV photon and 9 MeV、12 MeV electron of medical linear accelerator measured before the daily treatment in the whole year 2009,analyze the output stability of medical linear accelerator and the main influent factors,and take some countermeasures to assure the quality of radiation therapy.Results: The K-S tests of the output dose of 3 of the 4 energies(6 MeV,15 MeV,15 MV) are submitted to normal distribution(P 0.05).Because of the RF driver malfunction,the output dose of 6 MV increased continuously from 98.4% to 102.5%,between the first month to third month.The K-S tests of the output of 6 MV are not submitted to normal distribution(P0.05).Conclusions: The output stability of medical linear accelerator is important to assure the quality of radiation therapy.Based on the output dose measurement before daily treatment,the analysis of the linear accelerator output stability helps to decrease the system error and aprove the accuracy of therapy dose.