Development and Validation of Candidate Reference Methods for Determination of Serum Magnesium Based on Flame Atomic Absorption Spectrometry

王金英,王清涛,马怀安,李惠玲,徐国宾,徐静,贾慧敏,童清
DOI: https://doi.org/10.3969/j.issn.1006-1703.2009.02.012
2009-01-01
Abstract:To establish and validate a candidate reference method for determination of serum magnesium based on flame atomic absorption spectrometry(FAAS),the samples of serum magnesium diluted by 100 folds at the presence of HCl and La2O3 were measured by AA6800 atomic absorption spectrophotometer.The results showed that the intra-and inter-assay coefficients of variation were 0.41%(0.17%~0.89%) and 0.79%(0.63%~0.90%) respectively,and the total imprecision was 0.83%(0.69%~0.92%).The Bias of assay was lower than 1.5%.The recovery rates of low and high level were 100.31% and 99.70% respectively,the sensitivity of assay was 1.87×10-3 mmol/L and the limit of detection was 0.0023mmol/L.The applicable measurement range was 0.5~2.0mmol/L,and the expected uncertainty was 0.9859%.The method for determination of serum magnesium based on FAAS is very accurate and precise;it might be regarded as a candidate reference method for the determination of total magnesium in serum.
What problem does this paper attempt to address?