Optimization of Electrodes Displacement for Transthoracic Defibrillation: A Simulation Study on Spatial Configuration Pattern

LAI Da-kun,CAO Tao-tao,WU Xiao-mei,FANG Zu-xiang
2007-01-01
Abstract:In this paper, 12 novel electrode pairs in various spatial displacement combination patterns were developed and simulated with a detail finite element model of the human to further improve the efficacy of transthoracic electrical defibrillation. Simulation results of every electrode pairs were obtained and compared quantitatively to determine the best displacement under three different defibrillation criteria of the defibrillation threshold voltage (DFT), the efficiency fraction of delivered current reaching the heart (EFF), and the uniformity of current density distributions in the heart (UNIF). In addition, a suggested electrode pair locating in the same plane reported by prior literature was also simulated and then compared with the new achieved best displacement. These results show that the combination of the right anterior thorax and the left lateral waist displacement in spatial configuration is the least likely to damage the heart during successful defibrillation and indicate that it is more efficient than the suggested position in the same plane.