[Optimization of Electrode Location and Size on Simulation in Electric Field Distribution of Atrial Defibrillation].

Cong Wang,Shengjun Yang,Yi Zheng,Xiaomei Wu,Qunshan Wang,Daming Wei
2014-01-01
Abstract:A distributed simulation method of electric field based on the atrial defibrillation of the heart modeling and finite element solution is proposed in this study. In order to solve the problem that ordinary clinical trials could not measure the actual distribution of the defibrillation electric field in the heart accurately, this method provides a research tool for electrical defibrillation. A complete atrial anatomical structure in the heart model is used in the research, the finite element method is proceeded to solve; Three parameters: defibrillation threshold voltage, the high field strength rate and the defibrillation threshold energy are set to evaluate the effect of defibrillation. The heart electric field distributions of transvenous atrial defibrillation with different electrode locations or sizes are simulated. The simulation results and the reported results match fairly well, which initially verify the feasibility of this method.
What problem does this paper attempt to address?