Comparative Study of Two Kinds of CAD Softwares for Volumetric Measurement of Lung Nodules--a Phantom Study

李敏,龚建平,张伟,钱铭辉
DOI: https://doi.org/10.3969/j.issn.1000-0313.2013.01.007
2013-01-01
Abstract:Objective:To evaluate the accuracy and the consistency of CT phantom nodule volumetry measurement with pulmonary CT intellective CAD software and Lung VCAR sotware.Methods:Spherical nodules 2.5mm,5.0mm,10.0mm and 20.0mm in diameter were placed within a commercially available lung phantom and scanned by a 64-slice CT scanner.Nodule volume(V) was measured by the formula V=(4/3)πr3.Automated nodule volume measurements were performed by using the two above-mentioned CAD softwares independently.Results:The volume of all 16 nodules within the phantom could be measured and analyzed by the two softwares.The average relative errors of nodules 2.5mm,5.0mm,10.0mm and 20mm in diameter were 21.1%,-4.7%,-1.6% and-2.3% respectively with pulmonary CT intellective CAD software.The average relative errors of nodules 2.5mm,5.0mm,10.0mm and 20.0mm in diameter were 15.5%,3.1%,2.7% and 1.4% respectively with Lung VCAR software.There was no substantial difference in the volume estimations between the two CAD softwares by Bland Altman analysis.Conclusion:Although the accuracy of volumetry measurement with pulmonary CT intellective CAD software was lower than that of lung VCAR,but still there was high consistency between the two CAD softwares.