Evaluation of the volume of pulmonary nodules with LungCARE software

DING Qi-yong,YANG Lu,HUA Yan-qing
2008-01-01
Abstract:Objective To evaluate the agreement of volumetric measurement of pulmonary nodules with LungCARE software in Siemens 16 slices CT.Methods A total of 116 pulmonary nodules(diameter from3.8-21 mm) in 27 patients were measured with LungCARE software by 2 radiologists(A and B) and one of 2 radiologists measured twice(A and C).The limits of intra-and inter-observers agreement and 95% confidence interval(95% CI) were calculated using the Bland and Altman method.The difference between two radiologists was evaluated by paired t-test.Results For intra-observer data(A and C) agreement a lower limit of agreement of-4.198%(95%CI:-4.21 to-3.80%) and an upper limit of agreement of 4.042%(95%CI:4.03 to 4.05%) were found.Inter-observer data(A and B) agreement yielded very similar limits of agreement:from-3.785(95%CI:-3.08 to 3.77%) to 3.655%(95%CI:3.64 to 3.67%).The difference between Intra-and inter-observer agreement was shown no significance(P=0.175).Conclusion A good agreement of less than 5% is obtained in the volumetric measurements of pulmonary nodules with Siemens LungCARE software.But the practical measurement may be considered by some effect factors such as reformation parameters and breathing movement,etc.